AVW216 2.4 GHz Wireless 2-Channel Vibrating Wire Analyzer Module
Réduit considérablement le bruit du signal
Une radio incluse dans l'interface
météorologie : eau : energie : gaz & flux turbulents : infrastructure : terre :

Aperçu

The AVW216 is a vibrating wire analyzer module that includes an internal 2.4 GHz spread-spectrum radio for wireless communication in many countries worldwide. With this vibrating wire analyzer module, your data logger can measure vibrating wire strain gages, pressure transducers, piezometers, tiltmeters, crackmeters, and load cells. These sensors are used in a wide variety of structural, hydrological, and geotechnical applications because of their stability, accuracy, and durability.

NB: En raison de changements dans la législation dans l'Union européenne (UE), ce type de radio et d’autres produits ne sont plus disponibles à la vente en Europe depuis le 01/01/2015. Par conséquent l'achat de ce produit n'est pas recommandé pour vos projets de nouveaux réseaux en Europe.

Lire la suite

Avantages et caractéristiques

  • Fournit de meilleures mesures en réduisant sensiblement les lectures erronées causées par les sources de bruit
  • Interfaces deux capteurs à corde vibrante; plus de capteurs peuvent être connectés si un multiplexeur AM16/32B est utilisé
  • Diagnostics d’autocontrôle donne une information en continue sur l'état du capteur
  • Haute résolution inférieure à 0,001 Hz (le standard de l'industrie est de 0,1 Hz)
  • Low current drain
  • Fonctionnement sans fil à distance, grâce à une radio intégrée à l'interface
  • Interface deux mesures de température et de fréquence pour les capteurs à corde vibrante

Images

Description technique

The AVW216 uses vibrating wire spectral-analysis technology (VSPECT®). VSPECT® observes the incoming sensor signal, performs a Fourier transform and a spectral analysis (transforming the time series into individual sinusoidal components in the frequency spectrum), and determines the sensor frequency by identifying the largest signal in the acceptable range while filtering out environmental and electrical noise.

The AVW216 analyzer module also provides many self-checking diagnostics such as vibrating element signal strength, signal-to-noise ratio, vibrating element signal decay ratio, and incorrect signal response. These diagnostics can be running in the background to give continual feedback of the condition for each sensor.

The AVW216 typically transmits its data to an RF416 spread-spectrum radio that is connected to the data logger. The AVW216 can also be connected directly to the data logger.


Compatibilité

Veuillez noter : Ce qui suit montre des informations de compatibilité générales. Ce n'est pas une liste complète de tous les produits compatibles.

Centrale de mesure

Produits Compatibilité Note
CR1000 (obsolète)
CR1000X
CR200X (obsolète)
CR216X (obsolète)
CR3000 (obsolète)
CR5000 (obsolète) The CR5000 supports the SDI-12 mode only.
CR800 (obsolète)
CR850 (obsolète)
CR9000X (obsolète)

Equipement de montage

Produits Compatibilité Note
ENC10/12 If using the RS-232 port, the AVW216 must be mounted flat to fit in an ENC10/12.
ENC12/14 If using the RS-232 port, the AVW216 must be mounted flat to fit in an ENC12/14.
ENC14/16
ENC16/18
ENC24/30
ENC24/30S

Informations de compatibilité supplémentaires

Capteurs

Le module d'analyse de spectre AVW216 pour cordes vibrantes permet à nos centrales d'acquisition de mesurer les jauges de contrainte à corde vibrante, les capteurs de pression, les piézomètres, les inclinomètres, les fissuromètres, et les cellules de charge.


Communications

L'AVW216 est conçue en incluant une radio à étalement de spectre et saut de fréquence, qui transmet les données à d'autres interfaces AVW216 ou des modems radio RF416 et au modem RF432. L'AVW216 ne possède pas le mode transparent de communication. elle ne peut pas communiquer avec les anciens modem RF415.


Considerations sur les coffrets

L'AVW216 requiert un dessicant,un environnement sans condensation. Une enceinte Campbell Scientific est recommandée lorsque l'interface est sur ​​le terrain. L'AVW216 intègre un œillet de fixation pour son montage sur la platine du coffret.Considérations sur l'alimentation

L'AVW216 est généralement alimenté directement par la centrale de mesure.

Spécifications

-NOTE- Electrical specifications are valid over a -25° to +50°C range unless otherwise specified. Non-condensing environment required.
Operating Temperature Range -25° to +50°C
Internal Radio Frequency Range 2.450 to 2.482 GHz
Radio Power 50 mW
Number of Vibrating Wire Sensors Measured Up to 2 vibrating wire sensors can be connected to the analyzer module. Additional sensors can be measured by using an AM16/32-series multiplexer.
Power Requirements 9.6 to 16 Vdc
Analog Input/Outputs 2 differential (DF) vibrating wire measurements (V+ and V-) and 2 single-ended (SE) ratiometric resistive half-bridge measurements (T+ and T-) for vibrating wire sensor's onboard temperature sensor.
Digital Control Ports 3 digital control ports (C1 – C3)
  • C1 functions as an SDI-12 I/O communication port.
  • C2 functions as a Clk output for multiplexer control.
  • C3 functions as a Reset output for multiplexer control.
RS-232 Port 1 9-pin RS-232 port (for connecting to a data logger COM port)
Measurement Resolution 0.001 Hz RMS (±250 mV differential input range; -55° to +85°C)
Measurement Accuracy ±0.013% of reading (±250 mV differential input range; -55° to +85°C)
Input Voltage Range ±250 mV (differential) for vibrating wire inputs
Common Mode Range ±25 V
Baud Rates Selectable from 1200 to 38.4 kbps (ASCII protocol is one start bit, one stop bit, eight data bits, and no parity.)
Memory
  • Either 128 or 512 kB of SRAM
  • 2 MB of OS Flash
Dimensions 21.6 x 11.18 x 3.18 cm (8.5 x 4.5 x 1.2 in.)
Weight 0.43 kg (0.95 lb)

Typical Current Drain @ 12 Vdc

Quiescent, Radio Off ~0.3 mA
Radio Duty Cycling 1 s ~3 mA (includes quiescent current)
Radio Always On ~26 mA (radio transmit current 100 mA)
Active RS-232 Communication ~6 mA (3 s after communication stops, the current will drop to the quiescent current)
Measurement ~25 mA (averaged over the 2 s)

Téléchargements

AVW200 OS v.06 (551 KB) 10-06-2016

Current AVW200 firmware. Use the Device Configuration Utility version 1.13 or greater to send firmware and to configure the AVW200.

Historique des révisions

FAQ

Nombre de FAQ au sujet de(s) AVW216: 13

Développer toutRéduire tout

  1. Non. Ces modules ne peuvent pas fonctionner en tant que répéteur ou routeur; toutefois, les données peuvent être transférées au moyen d'un RF416 qui peut-être configuré comme répéteur/routeur.

  2. Non. Un AVW216 n'est pas capable de communiquer avec un autre module de la série AVW200 modules.

  3. The AVW216 uses a 50 mW radio, and the AVW206 and AVW211 use 100 mW radios. If a radio with a higher output is desired, consider using the RF450 with the AVW200. The RF450 transmits at 1 W, which is 10 times the power of the radios in the AVW devices.

  4. L'AVW200 doit-être connecté directement à une centrale de mesure via un câble.

    L'AVW216 intégre une radio à étalement de spectre et saut de fréquence, qui permet à une centrale d'acquisition de données de communiquer avec un dispositif AVW à distance via une liaison radio appropriée.

  5. Les modules de la série AVW200 ne sont pas nécessaires, mais ils réduisent les chances de lectures erronées en filtrant le bruit électrique parfois présent sur la mesure.

  6. Il faut 2 s pour que les modules de la série AVW200 acquièrent une mesure de chaque capteur. Un module de la série AVW200 connecté à deux multiplexeurs AM16/32B, chacun avec 16 capteurs à corde vibrante, nécessite environ 64 s pour acquérir les données.

  7. Deux causes probables sont les suivantes :

    • L'instruction SerialOpen() n'a pas été ajoutée dans le code de la centrale de mesure.
    • Le paramètre correct de l'adresse PakBus n'a pas été utilisé dans l'instruction AVW200.
  8. Des dizaines de modules de la série AVW200 peuvent être accessibles via une CR1000. Le facteur limitant est la fréquence d'échantillonnage à laquelle la CR1000 est réglée pour récupérer des données à partir de chaque module. Si la CR1000 a besoin de recueillir des données très fréquemment des modules, vous devez adapter le nombre de modules de la série AVW200 en fonction de la taille de votre programme et la vitesse d'échantillonnage, afin de garantir la collecte de données.