OBSOLETE ›
Ce produit n'est plus disponible et a été remplacé par: EE181. Certains accessoires, pièces de rechange ou des services peuvent être encore disponibles.
HMP45C-LQ Temperature and Relative Humidity Probe
Services disponibles
Réparation No
Etalonnage No
Support gratuit No

Aperçu

The HMP45C-LQ is a rugged, accurate temperature and relative humidity probe. The sensor measures relative humidity over the 0 to 100% RH range and temperature over the -40° to +60°C range.

Lire la suite

Avantages et caractéristiques

  • Bien adapté pour les applications à long terme, sans surveillance
  • Précis et robuste
  • Configured for use with the RAWS-P Fire Weather Station

Images

Description technique

For this version of the HMP45C, the cable terminates in a military style connector and has a user-specified lead length. The connector attaches to the RAWS-P enclosure, which mounts to a CM6/CM10 tripod, CM110-series tripod, or UT-series tower. The probe uses a HUMICAP capacitive polymer H chip to measure relative humidity over the 0 to 100% RH range. A PRT measures temperature over the -40° to +60°C range.


Compatibilité

The sensor cable terminates in a military style connector and a user-specified lead length. The connector attaches to the RAWS-P enclosure.

Mounting

The HMP45C-LQ should be housed in a 41003-5 10-plate naturally-aspirated solar radiation shield. This radiation shield mounts to a tripod or tower mast, crossarm, or user-supplied pole.

Spécifications

  • Supply Voltage: 12 Vdc nominal (typically powered by datalogger)
  • Current Consumption: ≤4 mA (active)

Relative Humidity

  • Sensor: Vaisala’s HUMICAP® H-chip
  • Measurement Range: 0.8% to 100% RH, non-condensing
  • Output Signal Range: 0.008 to 1 Vdc
  • Accuracy at 20°C:
    • Against Factory Reference: ±1% RH
    • Field-Calibrated Against References: ±2% (0% to 90% range);
      ±3% (90% to 100% range)
  • Temperature Dependence: ±0.05% RH/°C
  • Long-Term Stability: Typically, better than 1% RH per year
  • Response Time: 15 seconds with membrane filter (at 20°C, 90% response)
  • Settling Time: 500 milliseconds

Temperature

  • Temperature Sensor: 1000 ohm Platinum Resistance Thermometer
  • Measurement Range: -39.2° to +60°C
  • Output Signal Range: 0.008 to 1.0 V
  • Accuracy: see graph in HMP45C product brochure