The PVS5120C is a composite sampler that deposits its water samples into a nine-liter container. This sampler uses a vacuum sampler controller, the VSC100, to allow a Modbus RTU client, SDI-12 client device, simple pulse input, or Campbell Scientific PakBus data logger to control and communicate with the water sampler.
This sampler is lightweight, portable, and battery-powered. It can fit in a small-sized manhole and be suspended by a stainless-steel harness for sewer applications.
The PVS5120C can be used with a standard or large pump. The standard pump takes samples at 1.6 m/s (5.1 ft/s) for horizontal draws of 7.6 m (25 ft). The large pump takes samples at 2.2 m/s (7.1 ft/s) at 7.6 m (25 ft). Both of these velocities are typical values measured at sea level. Intake velocities will decrease as altitude increases.
Lire la suiteLe PVS5120C utilise une pompe à vide externe pour puiser de l'eau à travers un tube d'admission, au lieu de la pompe péristaltique classique qui induit l'écoulement en comprimant un tube flexible. Parce que le procédé sous vide perturbe moins les échantillons d'eau, il permet de mieux conserver la solution aqueuse d'origine, en particulier si la solution présente de fortes concentrations de solides en suspension.
Pour éviter toute contamination croisée, l'échantillonneur utilise la pression d'air (jusqu'à 28 psi ou 193,06 kilopascal) pour purger le tube de l'eau en excès. La pompe standard peut pomper l'eau aussi haut que la grande pompe et elle est tout à fait appropriée dans la plupart des applications, spécialement lorsque l'alimentation est une préoccupation. La grande pompe convient aux applications avec des distances de tuyaux plus longues ou plus élevés, car elle fournit une vitesse de pompage élevée.
Le PVS5120C peut s'interfacer avec les centrales de mesure de Campbell Scientific ou des centrales RTUs d'autres fabricants.
Nos centrales de mesure peuvent s'interfacer et communiquer avec la plupart des capteurs de turbidité, de niveau d'eau ou de capteur hydrologiques, ainsi que de contrôler l'échantillonneur à partir d'une horloge, d'un événement ou de conditions mesurées.
Specialized Applications | Fits in small-sized manholes. Can be suspended by stainless-steel harness for sewer applications. |
5/8 Inch ID Tubing Compatible | Yes |
Sample Container | One 9 L bottle |
Enclosure | LLDPE (linear low-density polyethylene), three-piece construction, and stainless-steel fittings |
Cooling System | Insulated container wall, cavity space for ice |
Horizontal Velocity | Refer to the "PVS5120 Intake Velocity versus Hose Length at Sea Level" plot. |
Storage Temperature Range | -20° to +50°C |
Operating Temperature Range |
|
Operating Voltage |
|
Body Case Diameter | 42.8 cm (16.85 in.) |
Height |
|
Weight |
|
Sampler with Standard Pump and No Battery | 10.4 kg (23 lb) |
Sampler with Large Pump and No Battery | 11.8 kg (26 lb) |
7 Ah Battery | 1.8 kg (4 lb) |
17 Ah Battery | 6.3 kg (14 lb) |
Controller |
|
Sampler Triggering Methods |
|
RS-485 Communication | Can return sampler status when triggering samples via pulse or time-based control. |
Backup Power Source | Internal lithium battery maintains program settings and information in case of power failure. |
Multi-purpose Button | Service sampler, collect manual sample, and auto deploy |
Auto Deploy | Automatically sets the purge and vacuum times based on hose length (between 25 and 100 ft) |
Sampler Status/Feedback | Bottle number, sample number, response code, sampler battery voltage (available after every sample) |
Vacuum System |
|
Pinch Valve | Fixed – normally open |
Purge Cycle | Adjustable from 1 to 800 s (280 s for SDI-12) |
Suction Cycle | Variable (until liquid contacts level electrode in metering chamber or adjusts automatically to two times the purge time) |
Horizontal Maximum Transport Distance | 67 m (220 ft) |
Metering Chamber Cover |
|
Volume Control Tube | 316 stainless steel |
Metering Chamber Level Electrode | 316 stainless steel |
Intake Hose | Ordered as a common accessory. Campbell Scientific offers PVC hose with 25-ft and user-specified lengths. Intake end can have a lead sinker or stainless-steel strainer. Sampler end can have a clamp or quick-connect termination. |
Discharge Hose Material |
|
Exemple de programmes CR1000 pour déclencher une PVS5120, qui a été configuré pour fonctionner à partir de commandes SDI 12, des commandes PakBus ou des commandes Modbus.
VSC100 Operating System.
Nombre de FAQ au sujet de(s) PVS5120C: 1
Lorsque l'élévation verticale (la distance verticale entre le niveau de la source d'eau et la pompe de l'échantillonneur d'eau) augmente, la vitesse du fluide à travers le système d'échantillonnage diminue. Par conséquent, une plus grande vitesse est nécessaire pour compenser une plus grande hauteur et pour préserver un échantillonnage plus représentatif de grandes particules et de solides en suspension. Pour maintenir la vitesse nécessaire, une pompe plus grande peut être requise.
En général, les affirmations suivantes sont vraies concernant la vitesse verticale :